This document describes the features and characteristics of Agilent Technologies, Inc. 2010 *G4240-90021*.
This Quick Start Guide outlines the necessary steps to check the HPLC-Chip/MS system before running the Agilent mAb-Glyco Chip. It includes an acquisition checklist covering hardware installation, such as the Chip Cube, injection loop, and capillary, as well as software requirements like MassHunter Analysis Software and the Glycan Database. The guide also specifies that the Deglycosylation Buffer must be installed on channel A1 and not older than three months.
| Part Number | G4240-64020 |
|---|---|
| Injection Loop Part Number | G1375-87315 |
| Capillary Part Number | G4240-87301 |
| Enablement Kit Part Number | G4240-64025 |
| Sample Injection Volume | 2 ul |
| Carryover Reduction Time | 5.20 minute(s) |
Ensure the Chip Cube is in backflush mode, the 8 uL injection loop is installed, and the yellow 25 pm x 105 cm PEEK coated fused silica capillary is installed between the Micro HiP Sampler and the Chip Cube.
Mass Hunter Qualitative Analysis software and the respective version of Personal Compound and Database Library Software must be installed.
The time required for Carryover Reduction is 5.20 minutes.