This document defines the procedure for the installation qualification, operation qualification and performance verification (IQ/OQ/PV) of Agilent 3200 series meters
This document serves as a Standard Operating Procedure (SOP) for validating Agilent 3200 series meters in regulated laboratory environments. It covers three main phases: Installation Qualification (IQ), Operation Qualification (OQ), and Performance Verification (PV). The guide ensures that the system is installed correctly, operates according to specifications, and maintains measurement accuracy.
The meter should display a temperature reading between 49.0 to 51.0 °C.
The meter should display a potential (mV) reading between -0.5 to 0.5 mV.
The meter should display a current between 680 to 720 nA.