61-681 Digital Light Meter Calibration Manual

This document describes the calibration procedure for the CHY Light Meter, including offset calibration and calibration under four standard illuminant sources.


61-681 Digital Light Meter Calibration Manual - cover page
Brand
61-681
Document type
Service Manual
Language
English
Pages
5
File format
PDF
File size
243 KB
Published
04 July, 2012
Updated
28 September, 2023
MD5 checksum
3D57FEC06D416AEEAD42942447D8A816

About this manual

The document outlines the offset calibration process, which involves turning off power, covering the sensor head, connecting a battery, and shorting specific pins before powering on to display 'CAL'.

Following the offset adjustment, the guide details how to calibrate the meter under four standard illuminant sources (A, B, C, and D) with values near 18.5, 185, 1850, and 18.5k Lux.

What's inside

  • Offset calibration
  • Standard illuminant source A
  • Standard illuminant source B
  • Standard illuminant source C
  • Standard illuminant source D
  • Calibration completion

Frequently asked questions

What are the standard Lux values for sources A, B, C, and D?

The standards used are CHY 18.86, 182.3, 1820, and 181.2k Lux respectively.

What should the LCD display show after offset calibration?

The LCD display should show 'A' once the meter is left on steadily.