61-310 Meter Field Test Adjustment Method

This document introduces the temperature test and adjustment method for the 61-310 meter. The meter uses a thermocouple to measure temperature, and the accuracy of the thermocouple ranges from 2% to 20% depending on the purity of the metals and construction. Although the accuracy of the thermocouple is not high, it can be compensated for by calibration under normal environmental conditions. This document introduces a method of using the freezing point and boiling point of water to calibrate the thermometer.


61-310 Meter Field Test Adjustment Method - cover page
Brand
61-310
Category
Other Electronics
Document type
Service Manual
Language
English
Pages
3
File format
PDF
File size
396 KB
Published
04 July, 2012
Updated
30 September, 2023
MD5 checksum
304D37285B1C398501706085040BA2AD

About this manual

This document provides a field test and adjustment method for the 61-310 meter, which uses thermocouples to measure temperature. It explains that low-cost instruments often have lower accuracy thermocouples but can be calibrated using standard reference points: water freezing at 0°C (32°F) and boiling at 100°C (212°F).

What's inside

  • Field Test and Adjustment Method
  • Thermocouple Accuracy
  • Calibration Standard Construction
  • Meter Disassembly Steps
  • Celsius Calibration Procedure
  • Fahrenheit Calibration Procedure
  • Re-assembly Instructions

Frequently asked questions

What temperature range should the meter display when calibrated for Celsius?

The meter should display 0 °C +/- 1°C when placed in slushy water and 100 °C +/- 12C when placed in boiling water.

What temperature range should the meter display when calibrated for Fahrenheit?

The meter should display 32 °F +/- 1°F in slushy water and 212 °F +/- 19°C in boiling water.

How should the screws be tightened when reassembling the meter?

You should not tighten the screws past slightly snug.