Agilent AA-975 Automated Multielement Analysis Application Note

This study describes an automatic flame atomic absorption system used for the sequential analysis of sodium, potassium, calcium, magnesium, aluminium, iron, zinc and manganese in Pinus Radiata after a nitric acid perchloric acid digestion


Agilent AA-975 Automated Multielement Analysis Application Note - cover page
Brand
AUTOMATED
Category
Other
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
347 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
E50511C75C2852DD80BF080521A275FA

About this manual

The document details a study using an automatic flame atomic absorption system to analyze sodium, potassium, calcium, magnesium, aluminium, iron, zinc, and manganese in Pinus Radiata. The analysis involved digesting plant samples with nitric and perchloric acid followed by sequential measurement using the AA-975 instrument connected to an HP-85 computer and PSC-55 autosampler.

The note provides specific instrumental conditions, including wavelengths, integration times, and gas flow rates for each of the eight elements. It outlines the sample preparation procedure, including oven drying and acid digestion, and presents results in tables showing concentration values and calibration graphics. The document also explains the calculation method for determining the percentage weight of elements in the original sample.

What's inside

  • Introduction
  • Instrumentation
  • Equipment and Reagents
  • Instrumental Conditions
  • Sample Preparation Procedure
  • Results
  • Calculation of percentage weight
  • Summary

Specifications

Elements AnalyzedNa, K, Ca, Mg, Mn, Zn, Fe, Al
Integration Time3.0 seconds (4.0 for Al)
Air Flow Rate13.0 L/min
Acetylene Flow Rate2.30 to 5.00 L/min
Nitrous Oxide Flow Rate11.0 L/min

Frequently asked questions

What elements were analyzed in the study?

The system was used for the sequential analysis of sodium, potassium, calcium, magnesium, aluminium, iron, zinc and manganese.

What is the integration time for most elements?

The integration time is set to 3.0 seconds for most elements, while 4.0 seconds is used for aluminium.

What is the air flow rate setup for the burner?

The air flow rate is set up to 13.0 litres per minute for all elements analyzed.