Agilent SpectrAA-20 Blood Lead Determination Application Note

This article describes a method for determining blood lead using a reduced-palladium matrix modifier in graphite furnace atomic absorption. The method allows for the use of ashing temperatures as high as 1000°C for blood lead determination. A low background absorbance of 0.060 enables the use of the more sensitive 217.0 nm lead line. Background correction can be easily performed using a deuterium lamp. The high dilution of blood in 0.2% Triton X-100 allows for short drying and ashing times. The use of the L'vov platform is not necessary at high ashing temperatures, as interferences are reduced and atomization is more effective from the tube wall. Carbon buildup does not occur. Zeeman background correction is not required. Accuracy has been established through comparison with a solvent extraction method and mean values from the UK External Quality Assessment Scheme.


Agilent SpectrAA-20 Blood Lead Determination Application Note - cover page
Brand
Blood
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
138 KB
Published
12 June, 2012
Updated
29 September, 2023
MD5 checksum
7D539CA1D1A6D128A2FB0D8230638D8D

About this manual

This document details a method for determining lead in blood using Graphite Furnace Atomic Absorption (GFAA). The method utilizes a reduced-palladium matrix modifier to allow high ashing temperatures (up to 1000 °C) and low background absorbance, enabling the use of the sensitive 217.0 nm lead line. The procedure was developed and validated using an Agilent SpectrAA-20 atomic absorption spectrometer and GTA-96 graphite tube atomizer.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Sample collection
  • Reagents
  • Method
  • Instrumental
  • Discussion
  • Instrumental Parameters
  • Furnace Parameters

Specifications

Wavelength217.0 nm
Slit width0.5 nm
Lamp current7 mA
Measurement time1.0 sec
Background correctionOn
Replicates3
Sample volume50 pL
Diluent concentration0.2% w/v
Modifier concentration500 mg/L

Frequently asked questions

What wavelength is used for lead determination?

The method uses the 217.0 nm lead line.

What is the recommended sample volume?

Normally 50 pL of blood is taken, though smaller volumes can be used.

What type of background correction is used?

Deuterium arc background correction is used.

What gases are used for the furnace?

Nitrogen is used for drying and atomization, while Argon is used for ashing.