MP-101/103 User Manual

This document introduces a pH and temperature measuring instrument with features such as waterproof microprocessor treatment, large LED display, and rubber protection ring. The instrument has automatic or manual temperature compensation, easy calibration method, low power indication, etc. The measurement range is 0.00 to 14.00 pH and 0 to 100℃, with an accuracy of 0.02 pH, 2mV, and 0.5℃, and a resolution of 0.01 pH, 1mV, and 0.1℃.


MP-101/103 User Manual - cover page
Brand
MP-101/103
Category
Other
Document type
User Manual
Language
Chinese
Pages
1
File format
PDF
File size
197 KB
Published
13 March, 2012
Updated
29 September, 2023
MD5 checksum
780BF3E18F1C4B6D7A669B801969E03A

About this manual

The MP-101/103 is a handheld pH/ORP meter with a waterproof housing, large LED display, and both automatic and manual temperature compensation. It runs on a DC 9 V battery and features auto power‑off after 10 minutes of inactivity.

The manual covers preparation steps, calibration using pH 4.00 and 7.00 solutions, switching between pH and ORP modes, and measurement procedures. It also lists dimensions, power‑indicator behavior, and basic operating instructions. This manual is written in Chinese.

What's inside

  • Specifications
  • Preparation
  • Calibration Procedure
  • Mode Switching
  • Measurement
  • Power Indicator

Specifications

Automatic temperature compensation rangeup to 100 °C (MP-103)
Manual temperature compensation rangeup to 100 °C (MP-101)
Calibration pointspH 4.00 and 7.00
Power supplyDC 9 V
Dimensions108 × 75 × 30 mm
Auto power‑off after inactivity10 minutes
pH measurement range0 – 14.00 pH

Frequently asked questions

How do I calibrate the MP-101/103?

Hold the CAL key for 3 seconds to enter calibration mode, then immerse the electrode in pH 4.00 solution until the display shows 7.00, and repeat the step with pH 7.00 solution.

What power source does the device use?

It operates on a DC 9 V battery.

What is the measurement range of the meter?

The meter measures pH from 0 to 14.00.

Download PDF (197 KB)