GeneQuant pro user manual
This manual introduces the GeneQuant pro UV/Vis spectrophotometer, covering unpacking, positioning, installation, safety labels, and the instrument’s front and rear panels. It explains required environmental conditions, power connection, and initial calibration before the instrument is ready for use.
The guide details operating procedures for nucleic acid, protein, and bacterial culture measurements, including method selection, sample handling, data output, and routine maintenance such as lamp replacement and cleaning. Technical specifications and compliance information are listed at the end of the document. This manual is written in Chinese.
| Wavelength range (fixed) | 230, 260, 280, 320, 546, 562, 595, 600 nm |
|---|---|
| Light source | Long‑life borosilicate glass pulsed deuterium lamp |
| Detector | Silicon photodiode |
| Power requirement | 100‑240 V AC ±10 %, 50/60 Hz, 80 VA |
| Dimensions | 320 × 400 × 160 mm |
Turn off the instrument, disconnect the power cable, remove the seven bottom screws, open the cover, detach the lamp connector, replace the lamp, re‑assemble, and reconnect power.
The instrument must be used indoors, at a temperature of 10 °C to 40 °C, with relative humidity up to 80 % at 31 °C, and placed on a stable, level bench that can support at least 4 kg.