This manual introduces the basic operation methods of CARDIAC SCIENCE AEDs, including AED overview, operating modes, batteries, electrodes, manual operation, etc.
About this manual
This manual provides instructions for safely using the Cardiac Science Powerheart AED G3 Pro. It covers the three operation modes—automatic, manual and ECG monitoring—along with safety symbols, self‑test procedures, and the RescueReady status indicator.
The guide also describes battery installation, charging, and specifications such as voltage, capacity, backup time, and lifespan. Contact information for service and technical support is included at the end of the document. This manual is written in Chinese.
What's inside
- AED Overview
- Operation Modes
- Battery Installation
- Manual Operation
- ECG Monitoring
- Safety Symbols
- RescueReady Indicator
- Energy Levels and Patient Impedance
- Contact Information
Specifications
| Battery voltage | 11.1 V |
|---|
| Battery chemistry | Lithium‑ion |
|---|
| Battery capacity | At least 60 shocks (standard 100) |
|---|
| Battery backup time | 6 months |
|---|
| Battery life | 2.5 years or 300 charge cycles |
|---|
| Battery weight | 1 lb 3 oz |
|---|
| Output voltage (max) | 12 V DC |
Frequently asked questions
How many shocks can the battery provide?
The battery can deliver at least 60 shocks (the standard is 100 shocks).
What does the RescueReady status indicator mean?
When the status indicator is green, the AED is in RescueReady state, meaning the self‑test has verified sufficient battery power, correctly connected electrodes, and proper internal circuitry.
How do I install the IntelliSense® battery?
Insert the battery as shown, push firmly until seated, then open the lid for 5 seconds; the status indicator will turn green.