Jinling GHX-010 Constant Temperature and Humidity Test Chamber Operation Manual
This is the Chinese operation manual (恒温恒湿试验箱操作说明书) from Ningbo Jinling Instrument Equipment Co., Ltd. for the GHX-010 test chamber. It introduces the company and the purpose of environmental simulation testing, then lists the technical parameters: a 1000×1000×1000 mm workspace, a temperature range of RT+10℃ to +200.0℃, a temperature fluctuation of ±0.5℃ and balanced temperature control (BTHC).
The manual then covers the chamber's construction and materials, the conditions required for installation (10℃~28℃ ambient, 86~106 Kpa, 220V 50Hz supply), operating precautions and safety notices. A large part describes the touch-screen temperature controller: the main screen, constant-value (FIX) operation with temperature and humidity setpoints, program (PROGRAM) operation, and program group setting, including parameter tables and screen figures. This manual is written in Chinese.
Covers: 产品概述 (Product Overview) · 技术参数 (Technical Parameters) · 产品结构与材质 (Structure and Material) · 产品使用条件 (Operating Conditions) · 产品结构 (Product Structure) · 使用注意事项 (Operating Precautions)
Latest Jinling manuals
Jinling GHX-010 Constant Temperature and Humidity Test Chamber Operation Manual
User Manual 19 pages · 1.8 MB · ZHChinese operation manual for the Jinling GHX-010 constant temperature and humidity test chamber: specifications, safety notes and controller operation.
Which document do I need?
User Manual
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in Jinling manuals
What is the temperature range of the GHX-010 test chamber?
The temperature range is RT+10℃ to +200.0℃, with a temperature fluctuation of ±0.5℃.
From: Jinling GHX-010 Constant Temperature and Humidity Test Chamber Operation Manual
What power supply does the chamber require?
The chamber requires a 220V 50Hz power supply.
From: Jinling GHX-010 Constant Temperature and Humidity Test Chamber Operation Manual