Agilent Rapid Screening of Melamine Cyanuric Acid in Milk Products Using Agilent J W HP-5ms GC Column Agilent 7890A/5975C GC/MSD with Column Backflushing Manual

Update: 30 September, 2023

A rapid screening method for the determination of melamine and cyanuric acid in milk products is developed using Agilent 7890A/5975C GC/MSD along with Agilent J&W HP-5ms GC columns. With the backflushing by capillary flow technologies, this method eliminates the time consumption for column bakeout after elution of target compounds, so as to significantly shorten the GC run time from more than 70 minutes to 14.5 minutes. Good linearity was obtained within the range of 10 to 200 µg/g, with correlation coefficients greater than 0.9996. The recoveries of both target compounds are greater than 95%.