Several oxidizable vitamins (vitamin A, B6, C, D3 and E) were analyzed using HPLC and electrochemical detection. Excellent sensitivity was achieved with all vitamins. The minimum detectable level for vitamin A- palmitate was about 80 pg, for β-carotene about 50 pg, for vitamin B6 about 30 pg, for vitamin C below 1 pg, for vitamin D3 about 70 pg and for vitamin E about 30 pg. The optimum potentials were evaluated using an automated increment for the potentials. The reproducibility was tested using vitamin B6 and was found to be 3% RSD of peak height for 100 runs. Due to the high selectivity of electrochemical detection, it was possible to keep sample preparation very simple.