Agilent GC- ECD Analysis Confirmation of Contract Laboratory Protocol Pesticides in Olive Oil Manual
Update: 30 September, 2023
This application note describes a method for analyzing 20 contract laboratory protocol (CLP) pesticides in olive oil using an Agilent GC-µECD system. A QuEChERS (Quick, Easy, Cheap, Effective, Rugged, and Safe) with dispersive solid phase extraction (dSPE) procedure cleaned the sample prior to analysis. A dual µECD and dual capillary GC column approach accomplished simultaneous primary and confirmatory analysis. The primary column, an Agilent J&W DB-35ms 30 m × 0.25 mm × 0.25 µm and a confirmatory column, an Agilent J&W DB-XLB 30 m × 0.25 mm × 0.50 µm effectively resolved all 20 CLP pesticides. An unpurged two-way capillary flow technology splitter divided the flow from a single injection port to the two analytical GC columns. Endosulfan sulfate and endosulfan 1 were present in the olive oil sample.