Featured Turner manuals

Turner BioSystems Modulus Light Meter Applications Note
The document describes a workflow that combines the Turner BioSystems Modulus multifunction light meter with the Quant‑iT DNA HS assay kit to provide precise quantification of double‑stranded DNA in small volumes. It outlines the required hardware, reagents, and step‑by‑step procedures for sample preparation, instrument setup, calibration, and analysis.
Key points include reagent dilution (1:200), avoidance of glass containers, calibration using five standard samples, and reading results in ng/µl on the instrument display. Contact information for both Molecular Probes, Inc. and Turner BioSystems is also provided. This manual is written in Chinese.
Covers: Product Description · Materials Required · Reagent Preparation · Instrument Setup · Calibration · Sample Analysis
Latest Turner manuals
Turner BioSystems Modulus Light Meter Applications Note
Application Note 3 pages · 111 KB · ZHThis applications note explains how to use the Turner BioSystems Modulus light meter with the Quant‑iT DNA HS kit for accurate dsDNA quantification.
Which document do I need?
Application Note
How to use the product for one specific task, with worked examples.
Questions answered in Turner manuals
What sample volume should be used for DNA quantification with the Modulus meter?
A sample volume of 210 µl is used, providing a linear range of 5‑100 ng of dsDNA.
From: Turner BioSystems Modulus Light Meter Applications Note
How should the Quant‑iT DNA HS reagent be prepared?
Dilute the concentrated dye 1:200 with Quant‑iT DNA HS buffer, use only plastic containers, and keep the solution protected from light.
From: Turner BioSystems Modulus Light Meter Applications Note