Renishaw XL-80 Manuals

ManualLib has 2 documents for the Renishaw XL-80 (Industrial Robot): 1 user manual. Every file can be read online or downloaded as a PDF for free.

User Manual 1

Key facts

Model
XL-80
Industry
Industrial
Documents
2
Languages
ZH
Last added
2000-01-01

Start here

User ManualRenishaw XL-80 Laser Interferometer System User Manual

Renishaw XL-80 Laser Interferometer System User Manual

User Manual 14 pages · 1.1 MB · ZH

This manual describes the XL‑80 laser head and the XC‑80 compensation unit, their mechanical dimensions, power requirements, and connection options. It also explains the supported software packages (LaserXL™ and QuickViewXL™) and how to configure the system for linear, angular, and flatness measurements.

Performance specifications, calibration procedures, accessories such as tripods and portable cases, and warranty information are covered. The document is intended for users who need portable, high‑accuracy measurement and calibration of machine tools and coordinate measuring machines. This manual is written in Chinese.

Read online

Renishaw XL-80 specifications

Laser head dimensions (L×W×H)214 mm × 120 mm × 70 mm
Laser head weight1.85 kg
Power input (laser head)90‑264 VAC
Linear measurement distance80 m
Linear measurement accuracy±0.5 ppm
Laser frequency stability±0.05 ppm (1 yr)
Maximum measurement speed4 m/s
Dynamic acquisition frequency10 Hz – 50 kHz
Pre‑heat time< 6 minutes
Operating temperature range0 °C – 40 °C
Storage temperature range-25 °C – 70 °C
Material temperature sensor accuracy±0.1 °C (0 °C – 55 °C)
Air temperature sensor accuracy±0.2 °C (0 °C – 40 °C)
Air pressure range650 – 1150 mbar ±1 mbar
Relative humidity range0 % – 95 % (non‑condensing ±6 % RH)

Source: Renishaw XL-80 Laser Interferometer System User Manual

All Renishaw XL-80 documents by type

User Manual 1

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What is the linear measurement accuracy of the XL‑80 system?

The XL‑80 provides a linear measurement accuracy of ±0.5 ppm across its full environmental range.

From: Renishaw XL-80 Laser Interferometer System User Manual

How long does the system take to warm up before measurement?

The system requires less than six minutes to pre‑heat before it can be used.

From: Renishaw XL-80 Laser Interferometer System User Manual