Wuhan特试特电气 Manuals

ManualLib has 2 Wuhan特试特电气 manuals, in categories such as Industrial Supply. Every file can be read online or downloaded as a PDF for free.

User Manual 2

Key facts

Manuals
2
Categories
1
Languages
ZH
Last added
2012-03-19

Latest Wuhan特试特电气 manuals

Which document do I need?

User Manual

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in Wuhan特试特电气 manuals

How should the TE‑OAT transformer be wired for an AC dielectric test?

Connect the low‑voltage input terminals a, x to the primary winding, use the measurement terminals E, F for the instrument, and connect the high‑voltage output terminals A, X to the test specimen as shown in Figure 6.

From: Wuhan Testyle TE-OAT/TE-ODT Transformer User Manual

How is the nominal transformer capacity calculated?

Use the formula Pn = k·Vn²·Ct·10⁻⁹, where Vn is the rated output voltage, Ct is the test specimen capacitance, and k is the safety factor based on Vn.

From: Wuhan Testyle TE-OAT/TE-ODT Transformer User Manual

What is the input voltage for the TE‑SBP generator?

The generator accepts a three‑phase 380 V, 50 Hz sinusoidal input.

From: TE TE-SBP Triple-Frequency High Voltage Generator Manual

How long can the generator run under load?

It can operate under load for 40 to 60 seconds.

From: TE TE-SBP Triple-Frequency High Voltage Generator Manual