Kaifeng Instrument XMUZ Level Display Installation Manual

The Kaifeng Instrument XMUZ level display manual provides installation, wiring, parameter setting, and maintenance instructions for XMUZ‑11/12/21/22 units.


Kaifeng Instrument XMUZ Level Display Installation Manual - cover page
Brand
开封仪表厂液位仪表分厂
Category
Industrial Supply
Document type
Installation Guide
Language
Chinese
Pages
8
File format
PDF
File size
226 KB
Published
17 October, 2011
Updated
28 September, 2026
MD5 checksum
12F65F59B7E96320FE419547A1AB91D6

About this manual

This manual describes the XMUZ digital level display instrument, which can be used with 4‑20 mA or RS‑485 level transmitters for monitoring liquid levels in chemical, petroleum, metallurgical, and hydro‑electric applications. It outlines the device's functions, such as digital and analog display, alarm limits, and remote data transmission up to 1.2 km.

The guide covers installation dimensions, wiring diagrams, power requirements, and step‑by‑step parameter configuration, including unit, range, upper/lower limits, and zero offset. Maintenance advice, storage conditions, and ordering information are also provided. This manual is written in Chinese.

What's inside

  • Function and Application
  • Technical Specifications
  • Operating Principle
  • Panel Layout
  • Installation and Wiring
  • Parameter Setting and Operation
  • Maintenance and Storage
  • Supplied Accessories
  • Ordering Information

Specifications

Input signal4‑20 mA DC, input impedance ≥1 MΩ, RS‑485 serial
Output signal (alarm relay)1 A/30 V DC, 0.3 A/125 V AC
Digital display4 digits, 1 s sample time, ±0.5 % FS error
Analog display error±1 % FS
Power supply220 VAC ±10 % 50 Hz ±5 %, power <10 W
Auxiliary DC outputs+24 VDC/100 mA, optional +12 VDC/100 mA
Dimensions150 mm × 160 mm × 80 mm
Weight≈0.5 kg
Operating temperature0 °C – 50 °C
Relative humidity≤85 %

Frequently asked questions

How do I set the measurement unit on the XMUZ display?

Press the ○ key to enter unit setting; the rightmost digit flashes. Enter 0 for cm or 1 for mm, then press ○ to confirm.

What temperature range can the instrument operate in?

The instrument works in an ambient temperature range of 0 °C to 50 °C.

What are the specifications of the alarm output contacts?

The alarm relay contacts are rated for 1 A at 30 V DC and 0.3 A at 125 V AC.

Download PDF (226 KB)